The nCPAP modes are designed so you only need to set the desired CPAP pressure. The flow is subsequently adjusted based on patient conditions and potential leaks. This prevents unintended peak pressures, guarantees highly efficient leak compensation, and helps to reduce oxygen consumption. Flow adjustment occurs very rapidly due to the high sensitivity of the pressure measurement.
Apnea is detected by MediTRIG. As a result, breaths are automatically triggered to stimulate breathing. Frequency, pressure level and breath duration can be individually adjusted.
In addition to the PEEP, breaths with fixed settings (pressure level, duration, frequency) are administered. These breaths have a higher pressure level and are not emitted in a synchronized manner.
Mode which is synchronized with the patient's spontaneous breathing with the aid of MediTRIG. With his/her inspiration, the patient can trigger an assisted breath which supports inspiration with flow and pressure.
nHFV combines the benefits of high-frequency ventilation and non-invasive CPAP support. The patient breathes spontaneously with CPAP support. A special valve shifts the air flow in all phases of spontaneous breathing under high-frequency oscillations (300 to 1200/minute).
The MediTRIG respiratory trigger was specially developed by medin for use in premature infants and newborns. Through the combination of the pressure measurement in the Medijet® generator and the signal processing in the medin nCPAP devices, MediTRIG works without additional sensors.
MediTRIG detects central as well as obstructive apnea.